Processing

NMRium can perform a Fourier transformation (FT) on the raw data of one-dimensional NMR spectra. To process an FID, drag it into the Drag and drop here box.

NMRium automatically performs zero filling, Fourier transformation, and phase correction. You can review every step in the Processings panel. Toggle the checkbox on a line to reversibly disable or re-enable a step. To remove a step permanently, click the recycle-bin icon on its line.

Fourier transformation

To prepare the FID for the Fourier transformation, click onto the Zero Filling icon to the left of the spectrum. Set the desired values for Size (zero filling) and Line Broadening (exponential window function) and click Apply. Then start the Fourier transformation by clicking on the FFT Filter button to the left of the spectrum.

Size and LB values

If you are unsure which values to set, you can use the following recommendations as a reference:

Size: Choose twice as many points as in the original FID.

Line Broadening (LB): Specify 0.3 Hz for ¹H spectra and 1-3 Hz for ¹³C.

NMRium has a live function. If you edit the raw data, you can see the changes directly. The changes are only saved when you click on the Apply button. If you want to hide the live function, uncheck the Live function above the workspace. This function is also available for many other NMRium functions.

Phase correction

Click the icon Phase correction to the left of the spectrum. You can choose between manual and automatic phase correction.

Automatic phase correction

Click the Apply button — the phase will be corrected automatically.

Manual phase correction

You can perform zero-order and first-order phase correction.

For zero-order phase correction, click and hold the green PH0 button above the spectrum. By moving the mouse to the left or right, you can perform the phase correction until the signal is correctly phased and release the button. By double-clicking, you can move the pivot point (red line) for the first-order phase correction.

Repeat the process for the first-order phase correction by pressing PH1 and moving the mouse. The focus should be on the signals that are furthest from the signal marked by the pivot point. Clicking the Apply button completes the phase correction.

Baseline correction

Click the Baseline correction icon to the left of the spectrum. You can choose between two algorithms: polynomial and airPLS.

  • For the polynomial algorithm, define a degree and click Apply.
  • For the airPLS algorithm, define a maximum iteration count and a tolerance, then click Apply.

The different processing steps are explained in the following sections: